Oxygen Analyser .
Oxygen Analyser
The
Oxygen Analyser measures oxygen concentrations from percentage levels
to a fraction of parts per million (ppm) in clean gases. Packaged on a
19â€Â rack mounting enclosure, it uses a oxygen sensor to provide
accurate measurement with minimal interference from other gases. The
unique cell and heater design gives very fast warm-up times, rapid
response, and long sensor life.
Specification of Oxygen Analyser
1. Display
- Multi-digit LCD, character height 12.7mm
2. Range
- 0.1ppm to 100% oxygen, autoranging
3. Display resolution
- From 10% to 100%: 0.1%
- From 0.5% to 9.99%: 0.01%
- From 500ppm to 4999ppm: 10ppm
- From 50ppm to 499ppm: 1ppm
- From 5ppm to 49.9ppm: 0.1ppm
- From 0ppm to 4.99ppm: 0.01ppm
4. Accuracy
- 100% to 100ppm: ±2% of reading or better
- 99ppm to 10ppm: ±1ppm
- 0.1ppm to 9/9ppm: ±0.2ppm
5. Stability
- Better than 2% of reading or 0.5ppm/month, whichever is greater
6. Sample flow
- Between 100 and 300ml/min for optimum operation
- Speed of response in clean inert atmospheres
- % levels: less than 0.5s for 90% change
- Levels from 100 to 10ppm: less than 5s for 90% change
- Levels less than 100ppm: less than 30s for 90% change
- When the cell is stabilised/conditioned at low levels, response to changes at that level is of the order of 3 to 4s
- Sample inlet pressure (no pump fitted)
- 10mbar minimum
- 8bar maximum
- Sample temperature (at analyser)
- 100°C maximum
- Sampling system materials
- Materials include: nickel, brass, aluminium, stainless steel, platinum, zirconia, alumina, PTFE and nitrile-rubber
- Sample connection (located on back panel of unit)
- Captive seal compression fitting suitable for 0.25inch (6mm) outside diameter tube
7. Analogue outputs
- Standard: 0 to 5V
- Optional: 4 to 20mA
- Each user-programmable between 0 – 100% to 0 – 5ppm
8. Alarm outputs
- 2 alarms: each user-configurable to OFF, HIGH or LOW status
- Hysteresis programmable from 0 to 10% of set-point
- Volt-free changeover relay contacts rated 48V ac or dc, 0.5A, normally-energised
- Ambient temperature: 45°C maximum
9. Power supply: 110/120V, 50/60Hz at 80VA max, 220/240V, 50/60Hz at 80VA max
10. Mounting: 19â€Â Rack-mounting enclosure 3U high and 270mm deep