DRY BLOCK CALIBRATOR .
Specification:

•    Range:     –25 °C to 140 °C (–13 °F to 284 °F) at 23 °C ambient
•    Accuracy :    ± 0.25 °C (holes greater than 1/4" [6.35 mm]: ± 1 °C)
•    Stability:     ± 0.02 °C at –25 °C ± 0.04 °C at 140 °C
•    Well-to-Well Uniformity:     ± 0.1 °C between similarly sized wells
•    Heating Times:     18 minutes from ambient to 140 °C
•    Cooling Times:     20 minutes from ambient to –25 °C
•    Stabilization Time:     7 minutes
•    Immersion Depth:     124 mm (4.875")
•    Inserts     Insert A, B, C, or D included (specify when ordering)
•    Outside Insert Dimensions     31.8 mm dia. x 124 mm (1.25 x 4.88 in)
•    Computer Interface     RS-232 included with free Interface-it software (Model 9930)
•    Power     115 V ac (± 10 %), 1.3 A or 230 V ac (± 10 %), 0.7 A, switchable, 50/60 Hz, 150 W
•    Size (W x H x D) : not more than 143 x 261 x 245 mm  (5.63 x 10.25 x 9.63 in)
•    Weight: not more than 5.7 kg (12 lb)
•    NIST-Traceable Certificate:     Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, and
140 °C