Technical Description of Throttling Calorimeter
Both very large and very small water contents of steam can be determined with precision by means of the two-stage process. For the direct determination of high water contents of x=0.5 and below the separating calorimeter is used. High vapour contents between approx. x=0.95 and x=1.0 can be determined indirectly using the throttling calorimeter. The vapour is adiabatically expanded and transferred to superheated steam. The low and high water contents can then be determined with the aid of the h-s diagram. Steam up to max. 10bar and 240°C can be investigated.
Specification of Throttling Calorimeter
Technical Data of Throttling Calorimeter
Steam
a. Max.: 240°C
b. Max.: 10bar
c. Connections: 1/2"
Measuring ranges
Dimensions and Weight of Throttling Calorimeter